Der medizinischer Auto-Adapter CAR100M mit Car/Truck Eingangsstecker.

CAR100M

100 Watt medical automobile power adapter

  • 넓은 입력 전압 범위 10-32V
  • 최대 출력 전력 110Watt
  • 환자 안전 최대 고려한 Galvanic isolation
  • 요청 시 개별 출력 커넥터
  • 자동차/트럭 입력 커넥터
  • 벽걸이 키트 선택

더 많은 정보를 찾고 계십니까?

Downloads CAR100M

CAR100M

Specification

Input
Voltage range10.00 - 32.00VDC
Standby currentTypical 45.00mA @ 10.00V input
Typical 35.00mA @ 32.00V input
Input fuse13.00A
ProtectionReverse polarity 
Over voltage protection (Transient suppression circuit)
Output
Voltage range16.00 - 24.00V nominal
Power

100.00W cont.

Initial set tolerance

±3% max.

Load regulation

-4% max.

Line regulation

±1% max.

Ripple & noise

< 0.12V pk-pk

Current limitConstant current
Protection

Short circuit
Over temperature protection
Over voltage protection

Environmental

CoolingConvection cooled
Temperature

Operating: -20°C to 40°C (40°C to 50°C, derating 35%)
Non-operating: -40°C to 75°C

Pressure & altitude

Operating: 1200hPa to 570hPa (max. 4600m above sea level)
Non-operating: 1200hPa to 115hPa (max. 15000m above sea level)

Humidity5 to 95% r.H., non condensing
General
Galvanic isolationIsolation voltage 2.00kV
Input connector

Cigarette lighter adapter

Output connector

Customer-specific

Efficiency

typical 88% at 100% load

MTBF

> 100000h @ 25°C ambient temperature and Gper MIL-HDBK217F Humidity

Green procurementWEEE, RoHS
IndicatorSingle green LED
Ingress protection rating

IEC60529, IP42

Regulatory approvalsEuropeCE
 United KingdomUKCA
 International

CB Report IEC 60601-1 3rd edition

 USA & CanadaUL60601-1
 Australia & New ZealandRCM
Electromagnetic emissionsEurope

EN 55011/55022 class B

 USA & CanadaFCC15 class B
 Medical electrical equipment

IEC60601-1-2

 Airborne equipment

RTCA/DO-160F, Section 21

Electromagnetic immunityESD immunity

IEC61000-4-2, 8kV contact/15kV air

 Electromagnetic field immunity

IEC61000-4-3, 80-2700MHz, 30V/m, modulation 80%/200Hz

 Electrical fast transient/burst immunity test

IEC61000-4-4, +/-2kV, 5/50ns

 Surge immunity test

IEC61000-4-5, +/-0.5kVsym, 1.2/50(8/20)μs

 Conducted Immunity

IEC61000-4-6, 0.15-80MHz, 10V, 80%, 1KHz, AM, with CDN

 ISO-Impulses

ISO7637-2, pulse 1 till 5, immunity test level III

 Medical electrical equipment

IEC60601-1-2

 UNECE vehicle regulationE13 ECE Reg. 10

- Other approvals on request

Housing dimensions (LxWxH)

165.00 x 64.00 x 38.00mm, without cables and connectors

Input cable length

240cm ±5%

Output cable length

100cm ±5%

Weight

approx. 420g, without cables, connectors and wall-mount kit


CAR100M is approved worldwide